SETUP AND OPERATION SECTION 2
Page 2-5
something is not shorting the output. Wiring is
in SECTION 4, SERIAL PORT PIN OUT.
4. Check the serial parameters on your PC.
They should be set to:
19200 baud
No par ity
8 data bits
1 stop
Make sure you have set the communications port on
your PC to the one the VTC-9F is plugged into.
5. Receiving a sign on message and can't enter
characters.
Check U9, pin 14 for at least -6 volts with the serial
cable connected to the PC . W hen it is near 0 volts,
the terminal or PC' s Tx line is not connected. When
you press a character on the terminal or P C, you
should see the voltage go positive on the
oscilloscope.
If all of this fails, call technical sup port listed in
SECTION 1.